CR:172B Class 2 SLM, Data Logging, Audio Recording, 1:1 & 1:3 Octave Bands
Environmental and Occupational Noise measurements.
Ideal for:
- Environmental noise impact assessments
- Occupational & Industrial Hygiene Noise Evaluations
- Hearing Protector Selection using HML or Octave Band Methods
- Tonal analysis using 1:3 Octave Band Filters
- Detailed analysis using audio recording
- Noise Ordinance & Community Noise Assessments
- General Noise Measurements
Measurements
- LXY, LXYMax, LXYMin
- LXeq, LCPeak, LZPeak, LAPeak, LCeq-LAeq, LXE, LAIeq
- Graph of Short LAeq, LCPeak
- Integrators 2 & 3: TWA, Dose %, Est Dose %
- Real-Time 1:1 Octave Bands (Graphical & Numeric)
- Real-Time 1:3 Octave Bands (Graphical & Numeric)
- Tonal Noise Detection in 1:3 Octave Bands
- NR & NC values & curves
- Leq LF (20Hz to 200Hz)
- Up to 28 Statistical Ln% Values
- Time History of LXYMax, LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq, LAVG, 1:1 Octave Bands, 1:3 Octave Bands
- Tonal Noise Detection in 1:3 Octave Bands
- Audio recording during measurement
- Time, date & duration of measurement
Where X=A, C & Z Frequency Weightings and Y=F, S & I Time Weightings. All frequency & time weightings are measured simultaneously. Time history stored by data logging versions only. 1:1 Octave Bands (A, B & C Versions), 1:3 Octave Bands (B & C Versions), Tonal noise detection (C Versions), NR & NC curves (A, B & C Versions)
Product Info
A Total Solution for Occupational and Environmental Noise Measurements: Optimus Green Sound Level Meters
Do you need an all-in-one solution for your noise measurements? The Optimus Green gives you all the noise data you could possibly need for both environmental and occupational noise.
By measuring every parameter you need simultaneously, the Optimus Green records everything in one measurement. There's no risk of choosing the wrong function or missing any data. All you need to do to get going is switch on, calibrate and start. It's that simple.
All of your measurement data displays on the large OLED screen. You'll be able to see straight away if there are any noise issues that need further investigation. You can easily navigate screens during measurements to see the data you need.
Whether it's environmental noise monitoring, detailed noise analysis using audio recording or workplace noise measurements, the Optimus Green has you covered.
Why Choose the Optimus Green?
- Backed by our 15 Year "No Quibble" Warranty for your peace of mind
- High Performance to Class 1 or Class 2 Standard
- Quick, simple and easy to operate
- Simultaneous measurement and data logging of all available parameters
- Measures Fast, Slow & Impulse Time Weightings
- dB(A), dB(C) and dB(Z) Frequency Weightings included
- Real-time 1:1 & 1:3 Octave Band Filters
- Tonal noise analysis
- VoiceTagTM voice note recording
- AuditStoreTM measurement verification
- Acoustic FingerprintTM audio triggering, recording & alerts during measurements for replay and analysis
- Compatible with CK:670 & CK:680 environmental noise monitoring kits for a total environmental noise monitoring solution
Specifications
Applicable Standards
IEC 61672-1:2002 Class 1 or Class 2 Group X
IEC 60651:2001 Type 1 I or Type 2 I
IEC 60804:2000 Type 1 or Type 2
IEC 61252:1993 Personal Sound Exposure Meters
ANSI S1.4 -1983 (R2006), ANSI S1.43 - 1997 (R2007)
ANSI S1.25:1991
1:1 & 1:3 Octave Band Filters to IEC 61260 & ANSI S1.11-2004
Microphone
Class 1 MK:224 pre-polarized, Class 2 MK:216 pre-polarized
Microphone Preamplifier
MV:200 Removable Preamplifier (All versions)
Total Measurement Range:
20dB to 140dB RMS Single Range
Noise Floor: <18dB(A) Class 1, <21dB(A) Class 2
Frequency Weightings
RMS & Peak : A, C, & Z Measured Simultaneously
Frequency Bands: 10 Octave Bands (31.5Hz to 16kHz, C & D variants)
Time Weightings
Fast, Slow & Impulse Measured Simultaneously
Display
High resolution OLED display. Ambient light sensor & illuminated keypad
Memory
4GB with 32GB factory fit option
AuditStore
Measurement verification data stored in secure memory
Time History Data Rates (Global settings)
10ms, 62.5ms, 125ms, 250ms, 1/2 sec, 1 sec, 2 sec (User selectable)
VoiceTag Audio Recording
Up to 30 seconds of audio notes with each measurement
Acoustic Fingerprint Audio Recording
Off, Manual, Threshold Triggered, Advanced Trigger
User options:
Studio Quality - 96kHz/32bit WAV format
Standard quality - 16kHz/16bit WAV format
Pre-Trigger function
Integrators
Three simultaneous "virtual" noise meters. Integrator 1 is preset to Q3 for Leq functions. Integrators 2 & 3 can be configured with the following:
Exchange Rate: 3, 4 or 5 dB
Threshold: 70dB to 120dB (1 dB steps)
Time Weighting: None or Slow
Criterion Level: 70dB to 120dB (1 dB steps)
Criterion Time: 1 to 12 hours in 1 hour steps
Integrator Quick settings
EU, OSHA HC & OSHA NC, OSHA HC & ACGIH,
MSHA HC & MSHA EC, Custom 1 & Custom 2
Ln Statistical Values
14 independent statistical Ln values calculated from 1/16th LAF
7 preset to L1.0, L5.0, L10.0, L50.0, L90.0, L95.0 & L99.0
7 user defined Ln values. CR:172C & CR:171C allow for an additional 14 Ln values with independent time and frequency weighting.
Measurement control
Measurement control with user selectable duration of manual, 1 min, 5 min, 10 min, 15 min,30 mins, 1 hour, Lden
Automatic Synchronisation & Repeat
Pause & Back Erase with user selectable duration
Dimensions
Size: 283mm x 65mm x 30mm
Weight: 300gms/10oz
Batteries
4 x AA Alkaline
Battery life
Typically 12 hours with Alkaline AA
Typically 20 hours with Lithium AA Non-Rechargeable
Battery life is dependent upon the battery type, quality & screen brightness
Connections
USB Type B to PC
AC & DC Output via ZL:174 (2 x Phono, 1m)
Multi-pin IO for external power via ZL:171 cable (2.1mm socket)
External Power: 5v-15v via MultiIO socket via ZL:171 cable (2.1mm socket)
Tripod Mount
1/4" Whitworth socket
Case
Material: High Impact ABS-PC with soft touch back & keypad
Environmental
Temperature: Operating -10°C to +50°C, Storage -20°C to +60°C
Humidity: Up to 95% RH Non Condensing
Electromagnetic performance
IEC 61672-1:2002 & IEC 61672-2:2003
Except where modified by EN 61000-6-1:2007 & EN 61000-6-1:2007
Language options
English, French, German, Spanish. Other language options may be available
Software Support
NoiseTools Download, Configuration & Analysis software supplied as standard. Compatible with Microsoft Windows XP, Vista, 7 & 8 (32bit & 64bit)
Measurement Functions
CR:1720 & CR:1710
Displayed Functions
LXY, LXYMax, LXYMin, LXeq, LCPeak, LAPeak, LZPeak, LCeq-LAeq, LXE, LAIeq
Graph of Short LAeq, LCPeak
Integrators 2 & 3: TWA, Dose%, Est Dose%
Measurement Run Time
Integrators 2 & 3: TWA, Dose%, Est Dose%
14 Statistical Ln values
Stored Functions
LXYMax & Time History of LXYMax
LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Time History of LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Integrators 2 & 3: LAVG , TWA. %Dose
Time History of LAVG
Ln Values: 14 independent statistical values
Audio recording during measurement
CR:172A & CR:171A
Displayed Functions
LXY, LXYMax, LXYMin, LXeq, LCPeak, LAPeak, LZPeak, LCeq-LAeq, LXE, LAIeq
Graph of Short LAeq, LCPeak
Integrators 2 & 3: TWA, Dose%, Est Dose%
Real-Time 1:1 Octave Bands (Graphical & Numeric)
NR & NC values & curves
Measurement Run Time
Integrators 2 & 3: TWA, Dose%, Est Dose%
14 Statistical Ln values
Stored Functions
LXYMax & Time History of LXYMax
LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Time History of LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Integrators 2 & 3: LAVG , TWA. %Dose
Time History of LAVG
1:1 Octave Bands:Overall Leq & Leq Time History for each band, NR & NC values & curves
Ln Values: 14 independent statistical values
Audio recording during measurement
CR:172B & CR:171B
Displayed Functions
LXY, LXYMax, LXYMin, LXeq, LCPeak, LAPeak, LZPeak, LCeq-LAeq, LXE, LAIeq
Graph of Short LAeq, LCPeak
Integrators 2 & 3: TWA, Dose%, Est Dose%
Real-Time 1:1 Octave Bands (Graphical & Numeric)
Real-Time 1:3 Octave Bands (Graphical & Numeric)
NR & NC values & curves
Measurement Run Time
Integrators 2 & 3: TWA, Dose%, Est Dose%
Leq LF (20Hz to 200Hz)
14 Statistical Ln values
Stored Functions
LXYMax & Time History of LXYMax
LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Time History of LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Integrators 2 & 3: LAVG , TWA. %Dose
Time History of LAVG
1:1 & 1:3 Octave Bands:Overall Leq & Leq Time History for each band, NR & NC values & curves
Ln Values: 14 independent statistical values
Audio recording during measurement
CR:172C & CR:171C
Displayed Functions
LXY, LXYMax, LXYMin, LXeq, LCPeak, LAPeak, LZPeak, LCeq-LAeq, LXE, LAIeq
Graph of Short LAeq, LCPeak
Integrators 2 & 3: TWA, Dose%, Est Dose%
Real-Time 1:1 Octave Bands (Graphical & Numeric)
Real-Time 1:3 Octave Bands (Graphical & Numeric)
Tonal Noise Detection in 1:3 Octave Bands
NR & NC values & curves
Measurement Run Time
Integrators 2 & 3: TWA, Dose%, Est Dose%
Leq LF (20Hz to 200Hz)
28 Statistical Ln values
Stored Functions
LXYMax & Time History of LXYMax
LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Time History of LAeq, LCeq, LZeq, LCPeak, LZPeak, LAPeak, LAIeq
Integrators 2 & 3: LAVG , TWA. %Dose
Time History of LAVG
1:1 & 1:3 Octave Bands:Overall Leq & Leq Time History for each band, NR & NC values & curves
Tonal Noise Detection in 1:3 Octave Bands
Ln Values: 28 independent statistical values
Audio recording during measurement
where x=A ,C ,Z; y= F, S, I
Other functions may calculated by the NoiseTools software and displayed on download.
Downloads
Delivery
We aim to Delivery products with you within 4 weeks of receipt of order. However, in exceptional circumstances delivery may be up to 6 weeks due to manufacturer lead times. Please inform us if you require fast delivery as we will do our utmost to expedite your order.
Quantity Orders
Please contact us directly to discuss quantity purchases and additional discounts. We will subsequently provide you with a formal quote with the applied discount, which will be valid for 30 days
Hire
Many of the products you have seen are available for hire, so feel free to contact us if you would prefer to hire this instrument.